In this study the bovine κ-, α- and β-caseins, α-lactalbumin and β-lactoglobulin by HPLC-UV using an RP column was detected by developed a simple and rapid method. Gradient elution was carried out at a 1 ml/ min flow rate, and a temperature of 25 o C, using a mixture of two solvents. Solvent A was acetonitrile: water: trifluoroacetic acid (100:900:1) and Solvent B was acetonitrile:water:trifluoroacetic acid (900:100:1). The effluent flow was monitored by a UV detector at 220 nm. Different chromatographic profiles were obtained for bovine milk proteins which was prepared with different methods for RP-HPLC. It was suggested that using of Solvent A and Solvent B during the sample preparation by addition into the sample before injection at given ratio was allowed the separation more clear and effective than the other sample preparation methods. By the current sample preparation technique and elution practice which was occurred in this study, it was achieved detection of bovine milk proteins was being simple, rapid and sensitive provide.